CVE-2017-13993CWE-428

CVE-2017-13993: fallo en i-SENS, Inc. SmartLog Diabetes Management…

Publicada el · Actualizada el

Aviso por correo cuando entre en el KEV, tenga PoC pública o suba de nota.

3Vexday Risk Score

Sin señal de explotación. Ningún artefacto público de explotación conocido hasta ahora.

ssvc Trackepss 1.8%
An Uncontrolled Search Path or Element issue was discovered in i-SENS SmartLog Diabetes Management Software, Version 2.4.0 and prior versions. An uncontrolled search path element vulnerability has been identified which could be exploited by placing a specially crafted DLL file in the search path. If the malicious DLL is loaded prior to the valid DLL, an attacker could execute arbitrary code on the system. This vulnerability does not affect the connected blood glucose monitor and would not impact delivery of therapy to the patient.