CVE-2019-25719: fallo de gravedad alta en Dräger Infinity Acute Care System
Dräger Infinity M540 VG4.1.1 Spoofing and DoS via Network Message Handling
Publicada el · Actualizada el
Seguir esta CVE+ Monitorear Infinity Acute Care System+ Monitorear Standalone Infinity M540 patient monitor
Aviso por correo cuando entre en el KEV, tenga PoC pública o suba de nota.
21Vexday Risk Score
Sin señal de explotación. Ningún artefacto público de explotación conocido hasta ahora.
ssvc Trackcvss 8.8epss 0.1%
Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower contain network message handling vulnerabilities that allow network-adjacent attackers to spoof or tamper with data and cause denial-of-service conditions. Attackers with access to an enabled Infinity network port or physical proximity to a wireless access point can modify device settings such as alarm states or alarm limits, and overwhelm the system with incoming data causing the device to reboot and lose network functionality.
CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:L/VI:L/VA:H/SC:N/SI:N/SA:N
Productos afectados
Dräger · Infinity Acute Care SystemDräger · Standalone Infinity M540 patient monitorCVEs relacionadas — Dräger Infinity Acute Care System