CVE-2023-0888mediumCWE-95

CVE-2023-0888: fallo de gravedad media en B. Braun Space Battery Pack SP with Wi-Fi

Authenticated eval injection in B. Braun Space Battery pack SP with Wi-Fi

Publicada el · Actualizada el

Aviso por correo cuando entre en el KEV, tenga PoC pública o suba de nota.

13Vexday Risk Score

Sin señal de explotación. Ningún artefacto público de explotación conocido hasta ahora.

ssvc Trackcvss 4.9epss 0.6%
An improper neutralization of directives in dynamically evaluated code vulnerability in the WiFi Battery embedded web server in versions L90/U70 and L92/U92 can be used to gain administrative access to the WiFi communication module. An authenticated user, having access to both the medical device WiFi network (such as a biomedical engineering staff member) and the specific B.Braun Battery Pack SP with WiFi web server credentials, could get administrative (root) access on the infusion pump communication module. This could be used as a vector to start further attacks
CVSS:3.1/AV:N/AC:L/PR:H/UI:N/S:U/C:N/I:H/A:N