← volver
CVE-2023-25931mediumCWE-620

Medtronic Micro Clinician & InterStim X Clinician App Password Reset Issue

13Vexday Risk Score

Sin señal de explotación. Ningún artefacto público de explotación conocido hasta ahora.

ssvc Trackcvss 6.4epss 0.3%
probabilidad de explotación
0.3%top 81% de las CVE
explotación observada
noninguna fuente lo reporta
Medtronic identified that the Pelvic Health clinician apps, which are installed on the Smart Programmer mobile device, have a password vulnerability that requires a security update to fix. Not updating could potentially result in unauthorized control of the clinician therapy application, which has greater control over therapy parameters than the patient app. Changes still cannot be made outside of the established therapy parameters of the programmer. For unauthorized access to occur, an individual would need physical access to the Smart Programmer.
CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:L/I:H/A:H