Spacelabs Healthcare Sentinel 10.5.x < 11.6.0 Unauthenticated RCE via .NET Remoting
28Vexday Risk Score
Sin señal de explotación. Ningún artefacto público de explotación conocido hasta ahora.
ssvc Trackcvss 9.2epss 0.7%
probabilidad de explotación
0.7%top 50% de las CVE
explotación observada
noninguna fuente lo reporta
Spacelabs Healthcare Sentinel versions 10.5.x and higher and 11.x.x before 11.6.0 contain an unauthenticated remote code execution vulnerability through a deprecated .NET Remoting HTTP channel exposed on port 8989 that allows attackers to perform arbitrary file read and write operations by supplying valid .NET URI endpoints. Attackers can write ASPX webshells to the IIS wwwroot directory to achieve unauthenticated remote code execution on the system. Port 8989 is not exposed in a default Sentinel installation; exploitation requires that the .NET Remoting port has been explicitly made network-accessible through deliberate configuration or network policy changes.
CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N
Productos afectados
Spacelabs Healthcare · SentinelReferencias
https://spacelabshealthcare.com/products/diagnostic-cardiology/connectivity/sentinelhttps://spacelabshealthcare.com/wp-content/uploads/2026/06/079-0273-00-RevA-Security-Advisory-Sentinel-.NET-Remoting-Vulnerability.pdfhttps://www.vulncheck.com/advisories/spacelabs-healthcare-sentinel-x-unauthenticated-rce-via-net-remoting