CVE-2014-5432CWE-592

CVE-2014-5432: falha em Baxter SIGMA Spectrum Infusion System

Publicada em · Atualizada em

Aviso por e-mail quando ela entrar no KEV, ganhar PoC pública ou subir de nota.

3Vexday Risk Score

Sem sinal de exploração. Nenhum artefato público de exploração conhecido até agora.

ssvc Trackepss 2.6%
Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 is remotely accessible via Port 22/SSH without authentication. A remote attacker may be able to make unauthorized configuration changes to the WBM, as well as issue commands to access account credentials and shared keys. Baxter asserts that this vulnerability only allows access to features and functionality on the WBM and that the SIGMA Spectrum infusion pump cannot be controlled from the WBM. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes.
CVEs relacionadas — Baxter SIGMA Spectrum Infusion System

No mesmo produto, das mais perigosas para as menos.