CVE-2018-10596: falha de alta gravidade em Medtronic 2090 CareLink Programmer
Medtronic 2090 Carelink Programmer Improper Restriction of Communication Channel to Intended Endpoints
Publicada em · Atualizada em
Aviso por e-mail quando ela entrar no KEV, ganhar PoC pública ou subir de nota.
21Vexday Risk Score
Sem sinal de exploração. Nenhum artefato público de exploração conhecido até agora.
ssvc Trackcvss 7.1epss 1.3%
Medtronic 2090 CareLink Programmer
uses a virtual private network connection to securely download updates. It does not verify it is still connected to this virtual private network before downloading updates. The affected products initially establish an encapsulated IP-based VPN connection to a Medtronic-hosted update network. Once the VPN is established, it makes a request to a HTTP (non-TLS) server across the VPN for updates, which responds and provides any available updates. The programmer-side (client) service responsible for this HTTP request does not check to ensure it is still connected to the VPN before making the HTTP request. Thus, an attacker could cause the VPN connection to terminate (through various methods and attack points) and intercept the HTTP request, responding with malicious updates via a man-in-the-middle attack. The affected products do not verify the origin or integrity of these updates, as it insufficiently relied on the security of the VPN. An attacker with remote network access to the programmer could influence these communications.
CVSS:3.1/AV:N/AC:H/PR:L/UI:R/S:U/C:H/I:H/A:H
CVEs relacionadas — Medtronic 2090 CareLink Programmer
No mesmo produto, das mais perigosas para as menos.
CVE-2018-5448MEDIUMMedtronic 2090 Carelink Programmer Relative Path TraversalEPSS 0.7%CVE-2018-5446MEDIUMMedtronic 2090 Carelink Programmer Storing Passwords in a Recoverable FormatEPSS 0.4%CVE-2018-18984MEDIUMMedtronic 9790, 2090 CareLink, and 29901 Encore Programmers Missing Encryption of Sensitive DataEPSS 0.3%