CVE-2019-25722highCWE-798

CVE-2019-25722: falha de alta gravidade em Dräger SC 6002XL

Dräger SC Monitoring Devices Hard-coded Credentials and DoS

Publicada em · Atualizada em

Aviso por e-mail quando ela entrar no KEV, ganhar PoC pública ou subir de nota.

21Vexday Risk Score

Sem sinal de exploração. Nenhum artefato público de exploração conhecido até agora.

ssvc Trackcvss 7.2epss 0.2%
Dräger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) contain hard-coded plaintext credentials in source code and a denial-of-service vulnerability that allows local and remote attackers to compromise device integrity across all software versions. A local attacker with direct device access can use the hard-coded credentials to access service and clinical accounts and alter device configuration, while a remote attacker can send malformed network packets to cause repeated device reboots, ultimately resulting in loss of network connectivity and disruption of patient monitoring.
CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:L/VI:L/VA:H/SC:N/SI:N/SA:N
CVEs relacionadas — Dräger SC 6002XL

No mesmo produto, das mais perigosas para as menos.