CVE-2019-6540
Medtronic Conexus Radio Frequency Telemetry Protocol Cleartext Transmission of Sensitive Information
The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement encryption. An attacker with adjacent short-range access to a target product can listen to communications, including the transmission of sensitive data.
CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N
Produtos afetados
Medtronic · Amplia CRT-DMedtronic · Brava CRT-DMedtronic · CareLink 2090 ProgrammerMedtronic · CareLink MonitorMedtronic · Claria CRT-DMedtronic · Compia CRT-DMedtronic · Concerto CRT-DMedtronic · Concerto II CRT-DMedtronic · Conexus Radio Frequency Telemetry ProtocolMedtronic · Consulta CRT-DMedtronic · Evera ICDMedtronic · Maximo II CRT-DMedtronic · Maximo II ICDMedtronic · Mirro ICDMedtronic · Mirro MRI ICDMedtronic · MyCareLink MonitorMedtronic · Nayamed ND ICDMedtronic · Primo ICDMedtronic · Protecta ICD, Protecta CRT-DMedtronic · Secura ICDMedtronic · Virtuoso ICDMedtronic · Virtuoso II ICDMedtronic · Visia AF ICDMedtronic · Viva CRT-DQuer saber se a sua infraestrutura está exposta a isto?