CVE-2020-12048
3Vexday Risk Score
Sem sinal de exploração. Nenhum artefato público de exploração conhecido até agora.
ssvc Trackepss 0.5%
probabilidade de exploração
0.5%top 61% das CVEs
exploração observada
nãonenhuma fonte reporta
Phoenix Hemodialysis Delivery System SW 3.36 and 3.40, The Phoenix Hemodialysis device does not support data-in-transit encryption (e.g., TLS/SSL) when transmitting treatment and prescription data on the network between the Phoenix system and the Exalis dialysis data management tool. An attacker with access to the network could observe sensitive treatment and prescription data sent between the Phoenix system and the Exalis tool.
Produtos afetados
n/a · Baxter Phoenix Hemodialysis Delivery System