The Continuous Glucose Monitoring Service's Record Access Control Point (RACP) write handler `memcpy`s the entire attacker-supplied ATT write value into a fixed 20-byte BSS buffer.
21Vexday Risk Score
Sem sinal de exploração. Nenhum artefato público de exploração conhecido até agora.
ssvc Trackcvss 8.7epss 0.2%
probabilidade de exploração
0.2%top 95% das CVEs
exploração observada
nãonenhuma fonte reporta
A buffer overflow in the Bluetooth Continuous Glucose
Monitoring Service (CGMS) Record Access Control Point (RACP) write handler
allows an authenticated BLE peer to overflow a 20-byte static buffer into
adjacent BSS memory. The exploitable impact cannot be predetermined - it
is entirely dependent on the linker-assigned BSS layout of the specific
firmware build, which may vary.
CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N
Produtos afetados
Nordic Semiconductor ASA · nRF Connect SDK