CVE-2020-25183: high-severity vulnerability in Medtronic Smart Model 25000 Patient Reader
Medtronic MyCareLink Smart Improper Authentication
Published · Updated
E-mail alert when it lands in KEV, gets a public PoC or its score goes up.
21Vexday Risk Score
No sign of exploitation. No public exploitation artifact known so far.
ssvc Trackcvss 8epss 0.8%
Medtronic MyCareLink Smart 25000 contains
an authentication protocol vulnerability where the method used to authenticate between the MCL Smart Patient Reader and the Medtronic MyCareLink Smart mobile app is vulnerable to bypass. This vulnerability enables an attacker to use another mobile device or malicious application on the patient’s smartphone to authenticate to the patient’s Medtronic Smart Reader, fooling the device into believing it is communicating with the original Medtronic smart phone application when executed within range of Bluetooth communication.
CVSS:3.1/AV:A/AC:L/PR:N/UI:R/S:U/C:H/I:H/A:H
Affected products
Medtronic · Smart Model 25000 Patient ReaderRelated CVEs — Medtronic Smart Model 25000 Patient Reader
In the same product, most dangerous first.