Contec Health CMS8000
21Vexday Risk Score
No sign of exploitation. No public exploitation artifact known so far.
ssvc Trackcvss 7.5epss 0.8%
exploitation probability
0.8%top 47% of all CVEs
observed exploitation
nono source reports it
The CMS800 device fails while attempting to parse malformed network data sent by a threat actor. A threat actor with network access can remotely issue a specially formatted UDP request that will cause the entire device to crash and require a physical reboot. A UDP broadcast request could be sent that causes a mass denial-of-service attack on all CME8000 devices connected to the same network.
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H
Affected products
Contec Health · CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor