CVE-2022-41627: medium-severity vulnerability in AliveCor KardiaMobile
Published · Updated
E-mail alert when it lands in KEV, gets a public PoC or its score goes up.
13Vexday Risk Score
No sign of exploitation. No public exploitation artifact known so far.
ssvc Trackcvss 4.8epss 0.1%
The physical IoT device of the AliveCor's KardiaMobile, a smartphone-based personal electrocardiogram (EKG) has no encryption for its data-over-sound protocols. Exploiting this vulnerability could allow an attacker to read patient EKG results or create a denial-of-service condition by emitting sounds at similar frequencies as the device, disrupting the smartphone microphone’s ability to accurately read the data. To carry out this attack, the attacker must be close (less than 5 feet) to pick up and emit sound waves.
CVSS:3.1/AV:P/AC:H/PR:N/UI:N/S:U/C:N/I:L/A:H
Affected products
AliveCor · KardiaMobile