CVE-2025-1204highCWE-912

CVE-2025-1204: high-severity vulnerability in Contec Health CMS8000 Patient Monitor

Published

E-mail alert when it lands in KEV, gets a public PoC or its score goes up.

21Vexday Risk Score

No sign of exploitation. No public exploitation artifact known so far.

ssvc Trackcvss 7.7epss 0.5%
The "update" binary in the firmware of the affected product sends attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function triggers if the 'C' button is pressed at a specific time during the boot process. If an attacker is able to control or impersonate this IP address, they could upload and overwrite files on the device.
CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:P/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N