← back
CVE-2026-66875highCWE-306

Mira Hormone Monitor, Mira Android App Missing authentication for critical function

21Vexday Risk Score

No sign of exploitation. No public exploitation artifact known so far.

ssvc Trackcvss 8.7epss 0.3%
exploitation probability
0.3%top 81% of all CVEs
observed exploitation
nono source reports it
In the Mira hormone monitor device firmware v1.7.1.47 build 01070147, a remote unauthenticated attacker within BLE range (approximately 10–30 meters) can silently rebind the device to an attacker-controlled account, extract stored hormone measurements in cleartext, cause a denial-of-service via malformed or undocumented command opcodes, and passively track the user via a static random BLE address that never rotates.
CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N