CVE-2014-5406
Hospira LifeCare PCA Infusion System
The Hospira LifeCare PCA Infusion System before 7.0 does not validate network traffic associated with sending a (1) drug library, (2) software update, or (3) configuration change, which allows remote attackers to modify settings or medication data via packets on the (a) TELNET, (b) HTTP, (c) HTTPS, or (d) UPNP port. NOTE: this issue might overlap CVE-2015-3459.
AV:N/AC:H/Au:N/C:C/I:C/A:C
Productos afectados
Hospira · LifeCare PCA Infusion System¿Quieres saber si tu infraestructura está expuesta a esto?
Hablar con TrueHacking →Referencias
https://github.com/cisagov/CSAF/blob/develop/csaf_files/OT/white/2015/icsa-15-125-01.jsonhttps://ics-cert.us-cert.gov/advisories/ICSA-15-125-01https://www.cisa.gov/news-events/ics-advisories/icsa-15-125-01https://xs-sniper.com/blog/2015/06/08/hospira-plum-a-infusion-pump-vulnerabilities/http://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/ucm446809.htm