CVE-2014-5406CWE-345

Hospira LifeCare PCA Infusion System

21Vexday Risk Score

No sign of exploitation. No public exploitation artifact known so far.

ssvc Trackcvss 7.6epss 1.2%
exploitation probability
1.2%top 34% of all CVEs
observed exploitation
nono source reports it
The Hospira LifeCare PCA Infusion System before 7.0 does not validate network traffic associated with sending a (1) drug library, (2) software update, or (3) configuration change, which allows remote attackers to modify settings or medication data via packets on the (a) TELNET, (b) HTTP, (c) HTTPS, or (d) UPNP port. NOTE: this issue might overlap CVE-2015-3459.
AV:N/AC:H/Au:N/C:C/I:C/A:C