CVE-2018-8868mediumCWE-749

CVE-2018-8868: fallo de gravedad media en Medtronic 24950 MyCareLink Monitor

Medtronic MyCareLink Patient Monitor Exposed Dangerous Method or Function

Publicada el · Actualizada el

Aviso por correo cuando entre en el KEV, tenga PoC pública o suba de nota.

13Vexday Risk Score

Sin señal de explotación. Ningún artefacto público de explotación conocido hasta ahora.

ssvc Trackcvss 6.2epss 0.3%
Medtronic 24950 MyCareLink Monitor and 24952 MyCareLink Monitor contains debug code meant to test the functionality of the monitor's communication interfaces, including the interface between the monitor and implantable cardiac device. An attacker with physical access to the device can exploit other vulnerabilities to access this debug functionality. This debug functionality provides the ability to read and write arbitrary memory values to implantable cardiac devices via inductive or short range wireless protocols. An attacker with close physical proximity to a target implantable cardiac device can use this debug functionality.
CVSS:3.1/AV:P/AC:H/PR:L/UI:N/S:C/C:H/I:L/A:L