CVE-2018-8868: falha de média gravidade em Medtronic 24950 MyCareLink Monitor
Medtronic MyCareLink Patient Monitor Exposed Dangerous Method or Function
Publicada em · Atualizada em
Aviso por e-mail quando ela entrar no KEV, ganhar PoC pública ou subir de nota.
13Vexday Risk Score
Sem sinal de exploração. Nenhum artefato público de exploração conhecido até agora.
ssvc Trackcvss 6.2epss 0.3%
Medtronic 24950 MyCareLink Monitor and 24952 MyCareLink Monitor contains debug code meant to test the functionality of the monitor's communication interfaces, including the interface between the monitor and implantable cardiac device. An attacker with physical access to the device can exploit other vulnerabilities to access this debug functionality. This debug functionality provides the ability to read and write arbitrary memory values to implantable cardiac devices via inductive or short range wireless protocols. An attacker with close physical proximity to a target implantable cardiac device can use this debug functionality.
CVSS:3.1/AV:P/AC:H/PR:L/UI:N/S:C/C:H/I:L/A:L
CVEs relacionadas — Medtronic 24950 MyCareLink Monitor
No mesmo produto, das mais perigosas para as menos.
CVE-2018-10622MEDIUMMedtronic MyCareLink Patient Monitor Network Credential WeaknessEPSS 0.4%CVE-2018-8870MEDIUMMedtronic MyCareLink Patient Monitor Use of Hard-coded PasswordEPSS 0.4%CVE-2018-10626MEDIUMMedtronic MyCareLink 24950 Patient Monitor Insufficient Verification of Data AuthenticityEPSS 0.4%