← volver
CVE-2024-48974criticalCWE-494

Life2000 Ventilator does not perform proper file integrity checks when adopting firmware updates

28Vexday Risk Score

Sin señal de explotación. Ningún artefacto público de explotación conocido hasta ahora.

ssvc Trackcvss 9.3epss 0.2%
probabilidad de explotación
0.2%top 95% de las CVE
explotación observada
noninguna fuente lo reporta
The ventilator does not perform proper file integrity checks when adopting firmware updates. This makes it possible for an attacker to force unauthorized changes to the device's configuration settings and/or compromise device functionality by pushing a compromised/illegitimate firmware file. This could disrupt the function of the device and/or cause unauthorized information disclosure.
CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H