CVE-2024-48974criticalCWE-494

CVE-2024-48974: critical vulnerability in Baxter Life2000 Ventilation System

Life2000 Ventilator does not perform proper file integrity checks when adopting firmware updates

Published · Updated

E-mail alert when it lands in KEV, gets a public PoC or its score goes up.

28Vexday Risk Score

No sign of exploitation. No public exploitation artifact known so far.

ssvc Trackcvss 9.3epss 0.2%
The ventilator does not perform proper file integrity checks when adopting firmware updates. This makes it possible for an attacker to force unauthorized changes to the device's configuration settings and/or compromise device functionality by pushing a compromised/illegitimate firmware file. This could disrupt the function of the device and/or cause unauthorized information disclosure.
CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H