CVE-2024-48974: critical vulnerability in Baxter Life2000 Ventilation System
Life2000 Ventilator does not perform proper file integrity checks when adopting firmware updates
Published · Updated
28Vexday Risk Score
No sign of exploitation. No public exploitation artifact known so far.
ssvc Trackcvss 9.3epss 0.2%
exploitation probability
0.2%top 96% of all CVEs
observed exploitation
nono source reports it
The ventilator does not perform proper file integrity checks when adopting firmware updates. This makes it possible for an attacker to force unauthorized changes to the device's configuration settings and/or compromise device functionality by pushing a compromised/illegitimate firmware file. This could disrupt the function of the device and/or cause unauthorized information disclosure.
CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H
Affected products
Baxter · Life2000 Ventilation SystemRelated CVEs — Baxter Life2000 Ventilation System
In the same product, most dangerous first.
CVE-2024-48966CRITICALLife2000 service tools for test and calibration do not support user authenticationEPSS 0.7%CVE-2024-48967CRITICALLife2000 ventilator and Service PC lack sufficient audit logging capabilitiesEPSS 0.6%CVE-2024-48973CRITICALDebug port on Life2000 Ventilator serial interface is enabled by defaultEPSS 0.2%CVE-2024-48971CRITICALClinician Password and Serial Number Clinician Password are hard-coded in Life2000 VentilatorEPSS 0.2%CVE-2024-48970CRITICALLife2000 Ventilator microcontroller lacks memory protectionEPSS 0.2%CVE-2024-9832CRITICALNo limit on failed login attempts with Clinician Password or Serial Number Clinician Password on Life2000 VentilatorEPSS 0.2%