← volver
CVE-2025-0626highexplotación observadaCWE-912

Hidden Functionality vulnerability in Contec Health CMS8000 Patient Monitor

43Vexday Risk Score

Prioriza la corrección. Ella explotación observada por VulnCheck.

ssvc Actcvss 7.7epss 1.1%
de la publicación al arma
Publicada en NVD30 ene
VulnCheck+355d
probabilidad de explotación
1.1%top 37% de las CVE
explotación observada
VulnCheck
The "monitor" binary in the firmware of the affected product attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function also enables the network interface of the device if it is disabled. The function is triggered by attempting to update the device from the user menu. This could serve as a backdoor to the device, and could lead to a malicious actor being able to upload and overwrite files on the device.
CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:P/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N