CVE-2025-0626: fallo de gravedad alta en Contec Health CMS8000 Patient Monitor
Hidden Functionality vulnerability in Contec Health CMS8000 Patient Monitor
Publicada el · Actualizada el
Aviso por correo cuando entre en el KEV, tenga PoC pública o suba de nota.
43Vexday Risk Score
Prioriza la corrección. Ella explotación observada por VulnCheck.
ssvc Actcvss 7.7epss 1.1%
de la publicación al arma
Publicada en NVD30 ene
VulnCheck+355d
The "monitor" binary in the firmware of the affected product attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function also enables the network interface of the device if it is disabled. The function is triggered by attempting to update the device from the user menu. This could serve as a backdoor to the device, and could lead to a malicious actor being able to upload and overwrite files on the device.
CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:P/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N
Productos afectados
Contec Health · CMS8000 Patient MonitorCVEs relacionadas — Contec Health CMS8000 Patient Monitor
En el mismo producto, de las más peligrosas a las menos.
Referencias
https://www.bleepingcomputer.com/news/security/backdoor-found-in-two-healthcare-patient-monitors-linked-to-ip-in-china/https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-030-01https://www.cisa.gov/resources-tools/resources/contec-cms8000-contains-backdoorhttps://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication