CVE-2025-0626: falha de alta gravidade em Contec Health CMS8000 Patient Monitor

Hidden Functionality vulnerability in Contec Health CMS8000 Patient Monitor

Publicada em · Atualizada em

Aviso por e-mail quando ela entrar no KEV, ganhar PoC pública ou subir de nota.

43Vexday Risk Score

Priorize a correção. Ela exploração observada pelo VulnCheck.

ssvc Actcvss 7.7epss 1.1%
da publicação à arma
Publicada no NVD30 de jan.
VulnCheck+355d
The "monitor" binary in the firmware of the affected product attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function also enables the network interface of the device if it is disabled. The function is triggered by attempting to update the device from the user menu. This could serve as a backdoor to the device, and could lead to a malicious actor being able to upload and overwrite files on the device.
CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:P/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N