CVE-2024-48966criticalCWE-306

CVE-2024-48966: falha crítica em Baxter Life2000 Ventilation System

Life2000 service tools for test and calibration do not support user authentication

Publicada em · Atualizada em

Aviso por e-mail quando ela entrar no KEV, ganhar PoC pública ou subir de nota.

28Vexday Risk Score

Sem sinal de exploração. Nenhum artefato público de exploração conhecido até agora.

ssvc Trackcvss 10epss 0.7%
The software tools used by service personnel to test & calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are installed could obtain diagnostic information through the test tool or manipulate the ventilator's settings and embedded software via the calibration tool, without having to authenticate to either tool. This could result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H