CVE-2024-48966: critical vulnerability in Baxter Life2000 Ventilation System

Life2000 service tools for test and calibration do not support user authentication

Published · Updated

E-mail alert when it lands in KEV, gets a public PoC or its score goes up.

28Vexday Risk Score

No sign of exploitation. No public exploitation artifact known so far.

ssvc Trackcvss 10epss 0.7%
The software tools used by service personnel to test & calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are installed could obtain diagnostic information through the test tool or manipulate the ventilator's settings and embedded software via the calibration tool, without having to authenticate to either tool. This could result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H