CVE-2025-0683
Exposure of Private Personal Information to an Unauthorized Actor vulnerability in Contec Health CMS8000 Patient Monitor
In its default configuration, Contec Health CMS8000 Patient Monitor transmits plain-text
patient data to a hard-coded public IP address when a patient is hooked
up to the monitor. This could lead to a leakage of confidential patient
data to any device with that IP address or an attacker in a
machine-in-the-middle scenario.
CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:N/VC:H/VI:N/VA:N/SC:N/SI:N/SA:N
Produtos afetados
Contec Health · CMS8000 Patient MonitorQuer saber se a sua infraestrutura está exposta a isto?
Falar com a TrueHacking →Referências
https://www.bleepingcomputer.com/news/security/backdoor-found-in-two-healthcare-patient-monitors-linked-to-ip-in-china/https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-030-01https://www.cisa.gov/resources-tools/resources/contec-cms8000-contains-backdoorhttps://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication