Mirion Medical EC2 Software NMIS BioDose Incorrect Permission Assignment for Critical Resource
21Vexday Risk Score
Sem sinal de exploração. Nenhum artefato público de exploração conhecido até agora.
ssvc Trackcvss 8.7epss 0.4%
probabilidade de exploração
0.4%top 64% das CVEs
exploração observada
nãonenhuma fonte reporta
NMIS/BioDose V22.02 and previous versions rely on a Microsoft SQL Server database. The SQL user account 'nmdbuser' and other created accounts by default have the sysadmin role. This can lead to remote code execution through the use of certain built-in stored procedures.
CVSS:4.0/AV:N/AC:L/AT:N/PR:L/UI:N/VC:H/VI:H/VA:L/SC:N/SI:N/SA:N
Produtos afetados
Mirion Medical · EC2 Software NMIS BioDose