CVE-2025-62575highCWE-732

CVE-2025-62575: high-severity vulnerability in Mirion Medical EC2 Software NMIS BioDose

Mirion Medical EC2 Software NMIS BioDose Incorrect Permission Assignment for Critical Resource

Published

E-mail alert when it lands in KEV, gets a public PoC or its score goes up.

21Vexday Risk Score

No sign of exploitation. No public exploitation artifact known so far.

ssvc Trackcvss 8.7epss 0.4%
NMIS/BioDose V22.02 and previous versions rely on a Microsoft SQL Server database. The SQL user account 'nmdbuser' and other created accounts by default have the sysadmin role. This can lead to remote code execution through the use of certain built-in stored procedures.
CVSS:4.0/AV:N/AC:L/AT:N/PR:L/UI:N/VC:H/VI:H/VA:L/SC:N/SI:N/SA:N