CVE-2015-3952CWE-312

CVE-2015-3952: vulnerability in Hospira Plum A+ Infusion System

Published · Updated

E-mail alert when it lands in KEV, gets a public PoC or its score goes up.

3Vexday Risk Score

No sign of exploitation. No public exploitation artifact known so far.

ssvc Trackepss 1.0%
Wireless keys are stored in plain text on Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.