CVE-2015-3953: vulnerability in Hospira Plum A+ Infusion System
Published · Updated
E-mail alert when it lands in KEV, gets a public PoC or its score goes up.
3Vexday Risk Score
No sign of exploitation. No public exploitation artifact known so far.
ssvc Trackepss 2.0%
Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
Affected products
Hospira · Plum A+3 Infusion SystemHospira · Plum A+ Infusion SystemHospira · Symbiq Infusion SystemRelated CVEs — Hospira Plum A+ Infusion System
In the same product, most dangerous first.