CVE-2025-3424: high-severity vulnerability in Philips IntelliSpace Portal
3.2.1 Arbitrary File Read in insecure .NET Remoting TCP Channel
Published · Updated
E-mail alert when it lands in KEV, gets a public PoC or its score goes up.
21Vexday Risk Score
No sign of exploitation. No public exploitation artifact known so far.
ssvc Trackcvss 7.7epss 0.2%
The IntelliSpace portal application utilizes .NET
Remoting for its functionality. The vulnerability arises from the exploitation
of port 755 through the "Object Marshalling" technique, which allows
an attacker to read internal files without any authentication. This is possible
by crafting specific .NET Remoting URLs derived from information enumerated in
the client-side configuration files.
This issue affects IntelliSpace Portal: 12 and prior.
CVSS:4.0/AV:A/AC:H/AT:N/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N/S:P/AU:Y/R:U/V:C/RE:M/U:Green
Affected products
Philips · IntelliSpace PortalRelated CVEs — Philips IntelliSpace Portal
In the same product, most dangerous first.