CVE-2025-3425: high-severity vulnerability in Philips IntelliSpace Portal
Unauthenticated Remote Code Execution via .NET Deserialization
Published · Updated
E-mail alert when it lands in KEV, gets a public PoC or its score goes up.
21Vexday Risk Score
No sign of exploitation. No public exploitation artifact known so far.
ssvc Trackcvss 7.3epss 0.3%
The IntelliSpace portal application utilizes .NET Remoting for its functionality. The vulnerability arises from the exploitation of port 755 through the deserialization vulnerability. After analyzing the configuration files, we observed that the server had set the TypeFilterLevel to Full which is dangerous as it can potentially lead to remote code execution using deserialization. This issue affects IntelliSpace Portal: 12 and prior.
CVSS:4.0/AV:A/AC:H/AT:N/PR:N/UI:A/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N/S:P/AU:Y/R:U/V:C/RE:M/U:Green
Affected products
Philips · IntelliSpace PortalRelated CVEs — Philips IntelliSpace Portal
In the same product, most dangerous first.