CVE-2025-14095: fallo de gravedad media en Radiometer Medical Aps ABL800 BASIC and ABL800…
Privilege boundary violation in Radiometer Products
Publicada el
Seguir esta CVE+ Monitorear ABL800 BASIC and ABL800 FLEX Analyzers+ Monitorear ABL90 FLEX and ABL90 FLEX PLUS Analyzers
Aviso por correo cuando entre en el KEV, tenga PoC pública o suba de nota.
13Vexday Risk Score
Sin señal de explotación. Ningún artefacto público de explotación conocido hasta ahora.
ssvc Trackcvss 6.8epss 0.2%
A "Privilege boundary violation" vulnerability is identified affecting multiple Radiometer Products. Exploitation of this vulnerability gives a user with physical access to the analyzer, the possibility to gain unauthorized access to functionalities outside the restricted environment. The vulnerability is due to weakness in the design of access control implementation in application software.
Other related CVE's are CVE-2025-14096 & CVE-2025-14097.
Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency.
Required configuration for Exposure:
Physical access to the analyzer is needed.
Temporary work Around:
Only authorized people can physically access the analyzer.
Permanent solution:
Local Radiometer representatives will contact all affected customers to discuss a permanent solution.
Exploit Status:
Researchers have provided working proof-of-concept. Radiometer is not aware of any publicly available exploit at the time of publication. Note:
CVSS score 6.8 when underlying OS is Windows 7 or Windows XP Operating systems and CVSS score 5.7 when underlying OS is Windows 8 or Windows 10 operating systems.
CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Productos afectados
Radiometer Medical Aps · ABL800 BASIC and ABL800 FLEX AnalyzersRadiometer Medical Aps · ABL90 FLEX and ABL90 FLEX PLUS AnalyzersRadiometer Medical Aps · ABL9 AnalyzersRadiometer Medical Aps · AQT90 FLEX AnalyzersCVEs relacionadas — Radiometer Medical Aps ABL800 BASIC and ABL800…
En el mismo producto, de las más peligrosas a las menos.
Referencias
https://www.radiometer.com/myradiometer