CVE-2025-14095: medium-severity vulnerability in Radiometer Medical Aps ABL800 BASIC and ABL800…
Privilege boundary violation in Radiometer Products
Published
Follow this CVE+ Monitor ABL800 BASIC and ABL800 FLEX Analyzers+ Monitor ABL90 FLEX and ABL90 FLEX PLUS Analyzers
E-mail alert when it lands in KEV, gets a public PoC or its score goes up.
13Vexday Risk Score
No sign of exploitation. No public exploitation artifact known so far.
ssvc Trackcvss 6.8epss 0.2%
A "Privilege boundary violation" vulnerability is identified affecting multiple Radiometer Products. Exploitation of this vulnerability gives a user with physical access to the analyzer, the possibility to gain unauthorized access to functionalities outside the restricted environment. The vulnerability is due to weakness in the design of access control implementation in application software.
Other related CVE's are CVE-2025-14096 & CVE-2025-14097.
Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency.
Required configuration for Exposure:
Physical access to the analyzer is needed.
Temporary work Around:
Only authorized people can physically access the analyzer.
Permanent solution:
Local Radiometer representatives will contact all affected customers to discuss a permanent solution.
Exploit Status:
Researchers have provided working proof-of-concept. Radiometer is not aware of any publicly available exploit at the time of publication. Note:
CVSS score 6.8 when underlying OS is Windows 7 or Windows XP Operating systems and CVSS score 5.7 when underlying OS is Windows 8 or Windows 10 operating systems.
CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Affected products
Radiometer Medical Aps · ABL800 BASIC and ABL800 FLEX AnalyzersRadiometer Medical Aps · ABL90 FLEX and ABL90 FLEX PLUS AnalyzersRadiometer Medical Aps · ABL9 AnalyzersRadiometer Medical Aps · AQT90 FLEX AnalyzersRelated CVEs — Radiometer Medical Aps ABL800 BASIC and ABL800…
In the same product, most dangerous first.
References
https://www.radiometer.com/myradiometer