CVE-2025-14095: falha de média gravidade em Radiometer Medical Aps ABL800 BASIC and ABL800…
Privilege boundary violation in Radiometer Products
Publicada em
Seguir esta CVE+ Monitorar ABL800 BASIC and ABL800 FLEX Analyzers+ Monitorar ABL90 FLEX and ABL90 FLEX PLUS Analyzers
Aviso por e-mail quando ela entrar no KEV, ganhar PoC pública ou subir de nota.
13Vexday Risk Score
Sem sinal de exploração. Nenhum artefato público de exploração conhecido até agora.
ssvc Trackcvss 6.8epss 0.2%
A "Privilege boundary violation" vulnerability is identified affecting multiple Radiometer Products. Exploitation of this vulnerability gives a user with physical access to the analyzer, the possibility to gain unauthorized access to functionalities outside the restricted environment. The vulnerability is due to weakness in the design of access control implementation in application software.
Other related CVE's are CVE-2025-14096 & CVE-2025-14097.
Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency.
Required configuration for Exposure:
Physical access to the analyzer is needed.
Temporary work Around:
Only authorized people can physically access the analyzer.
Permanent solution:
Local Radiometer representatives will contact all affected customers to discuss a permanent solution.
Exploit Status:
Researchers have provided working proof-of-concept. Radiometer is not aware of any publicly available exploit at the time of publication. Note:
CVSS score 6.8 when underlying OS is Windows 7 or Windows XP Operating systems and CVSS score 5.7 when underlying OS is Windows 8 or Windows 10 operating systems.
CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H