CVE-2024-48971: critical vulnerability in Baxter Life2000 Ventilation System
Clinician Password and Serial Number Clinician Password are hard-coded in Life2000 Ventilator
Published · Updated
E-mail alert when it lands in KEV, gets a public PoC or its score goes up.
28Vexday Risk Score
No sign of exploitation. No public exploitation artifact known so far.
ssvc Trackcvss 9.3epss 0.2%
The Clinician Password and Serial Number Clinician Password are hard-coded into the ventilator in plaintext form. This could allow an attacker to obtain the password off the ventilator and use it to gain unauthorized access to the device, with clinician privileges.
CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H
Affected products
Baxter · Life2000 Ventilation SystemRelated CVEs — Baxter Life2000 Ventilation System
In the same product, most dangerous first.
CVE-2024-48966CRITICALLife2000 service tools for test and calibration do not support user authenticationEPSS 0.7%CVE-2024-48967CRITICALLife2000 ventilator and Service PC lack sufficient audit logging capabilitiesEPSS 0.6%CVE-2024-48973CRITICALDebug port on Life2000 Ventilator serial interface is enabled by defaultEPSS 0.2%CVE-2024-48970CRITICALLife2000 Ventilator microcontroller lacks memory protectionEPSS 0.2%CVE-2024-9832CRITICALNo limit on failed login attempts with Clinician Password or Serial Number Clinician Password on Life2000 VentilatorEPSS 0.2%CVE-2024-48974CRITICALLife2000 Ventilator does not perform proper file integrity checks when adopting firmware updatesEPSS 0.2%